• Improved precision and drift of results with reference detector technology.
• Easy to use and powerful
• Designed for tough conditions and use in remote areas
• Remote support and simple maintenance
• 7†Color LCD, Touch-screen.
• Shortcut keypad, internal thermal Printer.
• Reagent open system, supports flow cell and cuvette mode.
• Analytical modes include Kinetic, fixed time, end-point.
• Real-time curve monitoring .
• Large storage 250 test programs.
• Powerful QC function supporting West guard Multi-rule, Levey hennings plot.
• Auto lamp sleeping mode.
• 8 standard filters, 2 optional filter.
• Multi-types result output.
Light Source : Tungsten Halogen lamp. 6V /10W.
Wavelengths : 340,405,500, 546, 578,620, 670 (nm) , 2 More optional filter.
Photometric Range : 0.000 – 2.700 Abs.
Wavelength Accuracy : ± 1nm.
Half Brand Width : ≤5nm.
Aspiration Volume : 200~6000µL.
Carryover : ≤1%.
Memory : 250 test Programs, 200000 test results. .
Interface : RS- 232, USB, LAN.
Flow Cell : 30µL.
Cuvette Mode : 12.5mm*12.mm cuvette Optional.
Pump : Sipper Pump.
Temperature Control : 250C, 300C, 370C, ± 0.50C , Ambient Temperature.
Display : 7†Color touch screen LCD, (800*480) Pixels.
Input : USB Mouse & Keyboard.
Output : Internal printer Or External Printer (Optional) .
Net Weight : 6.7Kg.
Dimension L x W x H : 350mmX362mmX180mm.
Power Supply : AC100V~240V,50Hz/60Hz.
Environmental requirements.
Ambient temperature (operating) : 100C~400C.
Relative humidity : ≤85%.
Transport and Storage Environment: .
Temperature : -200C~550C.
Humidity : ≤93%.
Atmospheric Pressure : 80kPa~106kPa.